Estimate the nucleic acid copy concentration in the original sample by inverting the positive partition fraction using a Poisson occupancy model

Assumes equal partition volumes, random distribution of target molecules and correct classification of positive and negative partitions.

Count partitions identified as positive rather than negative. A count of 0 is valid.

Number of partitions accepted for analysis: an integer from 1 to 100,000.

Enter the volume verified for the instrument and experimental conditions. 1 nL = 0.001 µL.

Include all dilution of the original sample into the reaction mixture. Use 1 if undiluted. The factor is applied to the result only 1 time.

Both methods give the same Poisson point estimate. The exact method numerically inverts the two binomial tail probabilities.

Results

Results update automatically when the required values are entered.

Method, applicability and sources

Calculation method

With k positive partitions, N total partitions, volume V (nL) per partition and total dilution factor D: negative fraction q = (N − k)/N; mean copies per partition λ = −ln(q). Reaction-mixture concentration = λ × 1000/V (copies/µL); original-sample concentration = reaction-mixture concentration × D. The 95% Wilson score or Clopper–Pearson exact binomial interval for the positive fraction is monotonically transformed using λ = −ln(1 − p), then scaled by volume and dilution factor. The calculation limit is 100,000 partitions; the exact method uses at most 64 bisection steps.

Worked example (hypothetical values)

Example: with 20,000 total partitions, 5,000 positive partitions, volume 1 nL and dilution factor 10, the positive fraction is 25%, λ ≈ 0.287682, reaction-mixture concentration ≈ 287.682 copies/µL and original-sample concentration ≈ 2,876.82 copies/µL.

Applicability and limitations

Single-target dPCR measurements with known, uniform partition volumes, independent random distribution of target molecules, and classified positive and negative partitions. Excludes misclassification of positive partitions, partition-volume variation, between-replicate differences, pipetting and dilution errors, multiple-target co-occurrence analysis, instrument-specific total error, experimental detection limits and diagnostic decisions. No finite point estimate is available when all partitions are positive.

Method and sources

Method: E883 v1.0(2026-11-14):Poisson MLE + Wilson/Clopper–Pearson 95%. Data recorded: 2026-11-14.