Membrane bubble point and capillary pressure
Calculate entry pressure and equivalent cylindrical diameter using the Young–Laplace equation for wetted pores
Assumes liquid-wetted cylindrical pores. Inverting a measurement does not determine pore-size distribution or membrane quality.
Results
—
Results update automatically when the required values are entered.
Method, applicability and sources
Calculation method
Young–Laplace cylindrical pore model: ΔP = 4γcosθ/d; inverse d = 4γcosθ/ΔP, with γ in N/m, d in m and ΔP in Pa. With inputs in mN/m and µm and output in kPa, the numeric relationship is ΔP(kPa) = 4×γ(mN/m)×cosθ/d(µm). Enter θ in degrees; it is converted to radians. Shape and instrument corrections are excluded.
Worked example (hypothetical values)
Example: surface tension 72 mN/m, angle 0° and diameter 1 µm give ΔP=4×72×1/1=288 kPa. Inverting 288 kPa gives an equivalent cylindrical diameter of 1 µm.
Applicability and limitations
Wetted membrane pores approximated as cylindrical capillaries with known surface tension and contact angle. Excludes contact angles ≥90°, unwetted pores, pore-shape corrections, gas flow and instrument corrections. Maximum diameter is cylindrical-equivalent; it does not establish nominal pore size, actual size distribution, filtration performance, integrity or medical-product acceptance.
Methods used
- Entry pressure from pore diameter (ΔP = 4γcosθ/d): scope γ,d >0; 0°≤θ<90°; cylindrical pore and static equilibrium; region global
- Equivalent cylindrical diameter from differential pressure (d = 4γcosθ/ΔP): scope γ,ΔP >0; 0°≤θ<90°; cylindrical pore and static equilibrium; region global
Method and sources
Method: ΔP = 4γcosθ/d (static capillary pressure in a wetted cylindrical pore). Data recorded: 2026-11-14.