Guard Bands and Measurement Uncertainty Acceptance Limits

Report an issue

Apply a selected guard-band decision rule to confirmed specification limits and expanded uncertainty.

Enter specification limits, uncertainty and measured value for the same physical quantity and in the same unit. No unit conversion is performed.

“Hold region” is this tool’s classification. Follow separately established procedures for final disposition of values in a hold region or outside the specification limits.

What to calculate

Use the same unit for the limits and uncertainty U.

Enter an already determined expanded uncertainty, not standard uncertainty uc.

Used only when “Custom multiplier” is selected as the decision rule.

If entered, shows whether the value is in the acceptance interval, a hold region within specification, or outside the specification limits.

Results

Results update automatically when the required values are entered.

Method, applicability and sources

Calculation method

For expanded uncertainty U in the same unit as the measured value: simple acceptance uses w=0; a U-width guard band uses w=U; a custom multiplier uses w=rU (r≥0). The lower acceptance limit is L+w and the upper is H−w. The acceptance interval includes both endpoints. Hold regions are [L,L+w) on the lower side and (H−w,H] on the upper side. With w=0, there are no hold regions within specification. If the guard bands overlap so that L+w>H−w, no acceptance interval exists.

Worked example (hypothetical values)

With lower limit 10, upper limit 20, expanded uncertainty U=0.2 and w=U, the acceptance interval is 10.2–19.8, the lower hold region is [10,10.2), and the upper hold region is (19.8,20], all in the same unit.

Applicability and limitations

A single scalar measurement with separately established specification limits and expanded uncertainty U for the same quantity and unit. Does not combine uncertainties, estimate a coverage factor, calculate measurement distributions or incorrect-decision probabilities, handle asymmetric uncertainty, select standards, determine certification compliance or issue conformity statements for calibration certificates.

Method and sources

Method: ILAC-G8:09/2019 §§4.1, 4.2.3, 5.2;JCGM 106:2012 §8.3.2. Data recorded: 2026-11-13.