Guard Bands and Measurement Uncertainty Acceptance Limits
Apply a selected guard-band decision rule to confirmed specification limits and expanded uncertainty.
Enter specification limits, uncertainty and measured value for the same physical quantity and in the same unit. No unit conversion is performed.
“Hold region” is this tool’s classification. Follow separately established procedures for final disposition of values in a hold region or outside the specification limits.
Results
—
- Guard-band width w (input unit)
- —
- Lower hold region
- —
- Upper hold region
- —
- Measured value location
- —
Results update automatically when the required values are entered.
Method, applicability and sources
Calculation method
For expanded uncertainty U in the same unit as the measured value: simple acceptance uses w=0; a U-width guard band uses w=U; a custom multiplier uses w=rU (r≥0). The lower acceptance limit is L+w and the upper is H−w. The acceptance interval includes both endpoints. Hold regions are [L,L+w) on the lower side and (H−w,H] on the upper side. With w=0, there are no hold regions within specification. If the guard bands overlap so that L+w>H−w, no acceptance interval exists.
Worked example (hypothetical values)
With lower limit 10, upper limit 20, expanded uncertainty U=0.2 and w=U, the acceptance interval is 10.2–19.8, the lower hold region is [10,10.2), and the upper hold region is (19.8,20], all in the same unit.
Applicability and limitations
A single scalar measurement with separately established specification limits and expanded uncertainty U for the same quantity and unit. Does not combine uncertainties, estimate a coverage factor, calculate measurement distributions or incorrect-decision probabilities, handle asymmetric uncertainty, select standards, determine certification compliance or issue conformity statements for calibration certificates.
Method and sources
Method: ILAC-G8:09/2019 §§4.1, 4.2.3, 5.2;JCGM 106:2012 §8.3.2. Data recorded: 2026-11-13.