Microscope Magnification, NA and Pixel Sampling Calculator
Check infinity-corrected objective specifications against imaging conditions to estimate resolution, digital field of view and specimen-space pixel spacing.
For an infinity-corrected objective used with a compatible tube lens and camera adapter.
Compares theoretical resolution and pixel sampling. It does not guarantee actual image resolution or field brightness.
Results
—µm
—×
- Specimen field of view (horizontal)
- —mm
- Specimen field of view (vertical)
- —mm
- Specimen-space pixel spacing (horizontal)
- —µm/px
- Specimen-space pixel spacing (vertical)
- —µm/px
- Pixels per Rayleigh distance (horizontal)
- —px
- Pixels per Rayleigh distance (vertical)
- —px
- Minimum adapter magnification for the 2-pixel criterion
- —×
- Objective image magnification with the actual tube lens
- —×
- Abbe theoretical resolution (reference)
- —µm
- Sensor pixel pitch (horizontal)
- —µm/px
- Sensor pixel pitch (vertical)
- —µm/px
- Total visual magnification
- —×
Results update automatically when the required values are entered.
Method, applicability and sources
Calculation method
Rayleigh resolution r (µm)=0.61×wavelength (nm)÷1000÷objective NA. Abbe criterion d (µm)=wavelength (nm)÷1000÷(2×NA). Actual objective magnification in an infinity-corrected system=design magnification×actual tube-lens focal length÷design tube-lens focal length. Camera magnification=actual objective magnification×adapter magnification. Specimen field of view (mm)=effective sensor dimension (mm)÷camera magnification. Sensor pitch (µm/px)=sensor dimension (mm)×1000÷effective pixel count. Specimen-space pixel spacing=sensor pitch÷camera magnification. Pixels per Rayleigh distance=r÷specimen-space pixel spacing. At least 2px in both directions is used as a simple sampling indicator. Required adapter magnification=2×max(horizontal, vertical sensor pitch)÷(r×actual objective magnification). This does not establish general Nyquist sufficiency or certify actual image quality.
Worked example (hypothetical values)
Example: design magnification 40×, NA0.65, wavelength 550nm, design and actual tube-lens focal lengths both 180mm, adapter 0.5×, effective sensor area 6.4mm×4.8mm and 1920×1440px give Rayleigh resolution approximately 0.516µm, camera magnification 20×, specimen field of view 0.32×0.24mm and specimen-space pixel spacing approximately 0.1667µm/px in both directions.
Applicability and limitations
Imaging systems with known compatibility between the infinity-corrected objective and tube lens, known design tube-lens focal length and known actual effective sensor area. Excludes calculating finite-conjugate objective magnification from mechanical tube length; cross-manufacturer correction compatibility; immersion-medium or coverslip suitability; condenser NA; corrections for aberrations, MTF, light level, vignetting or pixel binning; actual resolving performance; and diagnostic or quality judgments about specimens.
Methods used
- Rayleigh/Abbe diffraction limits (Rayleigh 0.61λ/NA; Abbe λ/(2NA)): scope Diffraction-limited reference values when objective NA and effective wavelength are known.; region Global
- Infinity-corrected imaging and 2-pixel sampling (M=f_tube/f_objective; 2 px per Rayleigh distance): scope Imaging systems with manufacturer-compatible tube lenses and projection adapters and a confirmed effective sensor area.; region Global
Method and sources
Method: Analytical formulas from manufacturer documentation (cross-checked 2026-11-14). Data recorded: 2026-11-14.