Microscope Magnification, NA and Pixel Sampling Calculator

Report an issue

Check infinity-corrected objective specifications against imaging conditions to estimate resolution, digital field of view and specimen-space pixel spacing.

For an infinity-corrected objective used with a compatible tube lens and camera adapter.

Compares theoretical resolution and pixel sampling. It does not guarantee actual image resolution or field brightness.

Magnification marked on the objective, specified for its design tube lens.

Use the objective NA specified for the intended immersion liquid or medium.

Enter a wavelength supported by the objective and imaging system.

The objective manufacturer’s reference focal length, not mechanical tube length.

The focal length of the tube lens actually used.

Enter 0.5 for a 0.5× adapter. This is separate from eyepiece magnification.

Enter the physical size of the active imaging area, not the nominal inch-format designation.

If entered, total visual magnification is also displayed, assuming the same imaging system.

Results

Results update automatically when the required values are entered.

Method, applicability and sources

Calculation method

Rayleigh resolution r (µm)=0.61×wavelength (nm)÷1000÷objective NA. Abbe criterion d (µm)=wavelength (nm)÷1000÷(2×NA). Actual objective magnification in an infinity-corrected system=design magnification×actual tube-lens focal length÷design tube-lens focal length. Camera magnification=actual objective magnification×adapter magnification. Specimen field of view (mm)=effective sensor dimension (mm)÷camera magnification. Sensor pitch (µm/px)=sensor dimension (mm)×1000÷effective pixel count. Specimen-space pixel spacing=sensor pitch÷camera magnification. Pixels per Rayleigh distance=r÷specimen-space pixel spacing. At least 2px in both directions is used as a simple sampling indicator. Required adapter magnification=2×max(horizontal, vertical sensor pitch)÷(r×actual objective magnification). This does not establish general Nyquist sufficiency or certify actual image quality.

Worked example (hypothetical values)

Example: design magnification 40×, NA0.65, wavelength 550nm, design and actual tube-lens focal lengths both 180mm, adapter 0.5×, effective sensor area 6.4mm×4.8mm and 1920×1440px give Rayleigh resolution approximately 0.516µm, camera magnification 20×, specimen field of view 0.32×0.24mm and specimen-space pixel spacing approximately 0.1667µm/px in both directions.

Applicability and limitations

Imaging systems with known compatibility between the infinity-corrected objective and tube lens, known design tube-lens focal length and known actual effective sensor area. Excludes calculating finite-conjugate objective magnification from mechanical tube length; cross-manufacturer correction compatibility; immersion-medium or coverslip suitability; condenser NA; corrections for aberrations, MTF, light level, vignetting or pixel binning; actual resolving performance; and diagnostic or quality judgments about specimens.

Methods used

  • Rayleigh/Abbe diffraction limits (Rayleigh 0.61λ/NA; Abbe λ/(2NA)): scope Diffraction-limited reference values when objective NA and effective wavelength are known.; region Global
  • Infinity-corrected imaging and 2-pixel sampling (M=f_tube/f_objective; 2 px per Rayleigh distance): scope Imaging systems with manufacturer-compatible tube lenses and projection adapters and a confirmed effective sensor area.; region Global

Method and sources

Method: Analytical formulas from manufacturer documentation (cross-checked 2026-11-14). Data recorded: 2026-11-14.